Romana Engleza
Login

Sonde pentru Electroconductivitate

Sonda cu doi electrozi E.Conductivity

  SI301             SI3013

SI 308T

SI 301
SI 3013
In-line E.Conductivity probe
Applications from 0 to 2 mS
SI 308T In-line E.Conductivity probe + Pt100
Applications for high purity water.

Sonda cu patru electrozi E.Conductivity

    SI311

    SZ312.4

SI 311 4-Electrode E.Conductivity probe + Pt100
Applications for immersion and in-line applications from 200 �s to 2000 mS. Suitable for microtrasmitter type 080310.
SZ 312.4 4-Electrode E.Conductivity probe + NTC 10K
Applications for immersion and in-line applications from 200 �s to 2000 mS. Suitable for
C 3645.

E.Conductivity probes with microtransmitter

    ST 311

  ST 31011

ST 311 4-Electrode Conductivity probe with microtransmitter
080310. For in-line and immersion applications.
This probe is an assembling of SI311 + 080310 microtransmitter, suitable for C7685 or C565.2 controllers. Connection by cables + connector type SZ9481 (10 mt) or SZ9483 (30 mt) .
ST 31011 4-Electrode Conductivity probe with microtransmitter 080310.
For in-line and immersion applications.
Suitable for C7685 or C565.2 controllers connected by SZ9481 (10 mt) or SZ9483 (30 mt)

 

Electrodeless Conductivity probes

SI 315

SI 315.1

SI 315, SI 315.1

Electrodeless Conductivity probe + Pt100
In-line and immersion applications.
Suitable for microtransmitter 080315 and C3655 controller from 200 �s
to 20 S.

     
    E.Conductivity probes with microtransmitter
For in-line and immersion applications.
This probe is an assembling of SI311 + 080310 microtransmitter, suitable for C7685 or C565.2 controllers. Connection by cables + connector type SZ9481 (10 mt) or SZ9483 (30 mt) .
     
    Submersible probe
Principle of operation
When the electrodeless Conductivity sensor is immersed in the solution to be measured a conductive loop is created through the two toroidally wound coils.
An alternating Current is applied to one of the coils which induces a Current in the conductive loop.
The second coil is used to measure the Conductivity which is proportional to the induced Current in the solution.
The advantages of the electrodeless method are more apparent in measurement applications in which electrodes contamination and polarization of a conventional Conductivity system can lead to erroneous readings.

Probe assembly
The submersible probe is a 5 part assembly:
- a sensing toroidal element with a built-in Temperature sensor
- a mounting adapter screwed to the back of the sensing element.
The back end is FNPT threaded for 1" pipe mounting.
- a mounting adapter screwed to the bottom of the microtransmitter.
The back end is FNPT threaded for 1" pipe mounting.
- a microtransmitter type 080315
- an extension pipe threaded two ends MNPT,
not included in the package
This probe is compatible with C7685 and C565.2
B&C Electronics controllers..

Accessories
SZ 9481 10 mt. cable with connector 2231520
SZ 9483 30 mt. cable with connector 2231520
2231520 IP67 connector for cable

 

Special E.Conductivity probes

  SZ 3320.1
  SZ 3330.1

  SAN 621-3-1-58

SZ 3320.1
SZ 3330.1
For high Temperature/Pressure.
Applications High purity water.
SAN 621-3-1-58 Pressurizable/Sterilizable.
Applications High purity water, pharmaceutical industry.

  SZ 3300.1

  SZ 3300.1 Graphite Electrodes.
Applications From 200 �S to 20